±5.2˚C
±3.0˚C
±(0.05% of rdg +2.0˚C)
±(0.05% of rdg +2.0˚C)
±3.5˚C
±(0.05% of rdg +2.0˚C)
±(0.05% of rdg +2.0˚C)
±(0.05% of rdg +1.0˚C)
±(0.05% of rdg +2.0˚C)
±(0.05% of rdg +1.0˚C)
±(0.1% of rdg +1.5˚C)
±(0.1% of rdg +0.5˚C)
±2.7˚C
±1.7˚C
±(0.05% of rdg +1.0˚C)
±(0.1% of rdg +1.0˚C)
±(0.1% of rdg +1.5˚C)
±0.5˚C
-100<TS≤800
600<TS≤1820
-100<TS≤100
100<TS≤1100
-100<TS≤1370
B
K
100<TS≤300
R: 300<TS≤1600
S: 300<TS≤1760
400≤TS≤600
(23°C±5°C)
- When 30 minutes
or more have elapsed
after power was
switched on
- Sampling 1s/10ch
Measurement
accuracy
Measurement
accuracy *1
A/D converter
Temperature coefficient
Maximum input voltage
Withstand voltage
Common mode rejection ratio
Noise
0≤TS≤1300
0≤TS≤2000
E
T
J
N
- GND connected
Type
Number of input channels
Method
Maximum sampling speed
R/S
0≤TS≤100
- Filter ON (10)
Measurement Temperature Rang
Measurement Accuracy
Reference contact compensation accuracy
*1: Thermocouple diameters T: 0.32 Φ, others: 0.65 Φ
W
-200≤TS≤-100
-200≤TS≤-100
-200≤TS≤-100
-100<TS≤400
-200≤TS≤-100
Temperature
Humidity
Voltage
Photo MOS relay scanning system,
all channels isolated,
balanced input
10ms/1ch
M3 screw type, 10 channels
Thermocouple
20m/50m/100m/200m/500m
1/2/5/10/20/50/1-5V F.S.
K、J、E、T、R、S、B、N、W(WRe5-26)
0 to 100% (voltage 0 V to 1 V scaling conversion)
*with B-530 (option)
Voltage
16-bit Delta-Sigma A/D converter
(Effective resolution: approx. 1/40,000 of ± range)
Gain : 0.01% of F.S./°C
Zero : 0.02% of F.S./°C
Occurs when sampling speed is 10, 20, or 50 ms.
Between +/– terminals : 60Vp-p
Between input terminal/input terminal
Between input terminal/GND
Between input terminal/input terminal
Between input terminal/GND
: 60Vp-p
: 60Vp-p
: 1 minute at 350Vp-p
: 1 minute at 350Vp-p
At least 90 dB (50/60 Hz; signal source 300Ω or less)
At least 48 dB (with +/- terminals shorted)
±0.1% of F.S.
DescriptionItem
Input Unit Specifications
19