EasyManua.ls Logo

GRAPHTEC midi LOGGER GL220 - Input Unit Specifications

GRAPHTEC midi LOGGER GL220
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
±5.2˚C
±3.0˚C
±0.05% of rdg +2.0˚C
±0.05% of rdg +2.0˚C
±3.5˚C
±0.05% of rdg +2.0˚C
±0.05% of rdg +2.0˚C
±0.05% of rdg +1.0˚C
±0.05% of rdg +2.0˚C
±0.05% of rdg +1.0˚C
±0.1% of rdg +1.5˚C
±0.1% of rdg +0.5˚C
±2.7˚C
±1.7˚C
±0.05% of rdg +1.0˚C
±0.1% of rdg +1.0˚C
±0.1% of rdg +1.5˚C
±0.5˚C
-100<TS800
600<TS1820
-100<TS100
100<TS1100
-100<TS1370
100<TS300
R: 300<TS1600
S: 300<TS1760
400TS600
(23°C±5°C)
- When 30 minutes
or more have elapsed
after power was
switched on
- Sampling 1s/10ch
Measurement
accuracy
Measurement
accuracy *1
A/D converter
Temperature coefficient
Maximum input voltage
Withstand voltage
Common mode rejection ratio
Noise
0TS1300
0TS2000
- GND connected
Type
Number of input channels
Method
Maximum sampling speed
/
0TS100
- Filter ON (10)
Measurement Temperature Rang
Measurement Accuracy
Reference contact compensation accuracy
*1: Thermocouple diameters T: 0.32 Φ, others: 0.65 Φ
-200TS-100
-200TS-100
-200TS-100
-100<TS400
-200TS≤-100
Temperature
Humidity
Voltage
Photo MOS relay scanning system,
all channels isolated,
balanced input
10ms/1ch
M3 screw type, 10 channels
Thermocouple
20m/50m/100m/200m/500m
1/2/5/10/20/50/1-5V F.S.
KJETRSBNWWRe5-26
0 to 100% (voltage 0 V to 1 V scaling conversion)
*with B-530 (option)
Voltage
16-bit Delta-Sigma A/D converter
(Effective resolution: approx. 1/40,000 of ± range)
Gain : 0.01% of F.S./°C
Zero : 0.02% of F.S./°C
Occurs when sampling speed is 10, 20, or 50 ms.
Between +/– terminals : 60Vp-p
Between input terminal/input terminal
Between input terminal/GND
Between input terminal/input terminal
Between input terminal/GND
: 60Vp-p
: 60Vp-p
: 1 minute at 350Vp-p
: 1 minute at 350Vp-p
At least 90 dB (50/60 Hz; signal source 300Ω or less)
At least 48 dB (with +/- terminals shorted)
   ±0.1% of F.S.
DescriptionItem
Input Unit Specifications
19

Other manuals for GRAPHTEC midi LOGGER GL220