CCT10/CCT20 USER MANUAL
10
S P E E d 0 :low speed (factory default), 7.5 Hz
S P E E d 1 :standard speed, 15 Hz
S P E E d 2 :High speed, 30 Hz
Press 【M+】key to confirm, and press 【CE】key to
exit.
UF-8
Display condition
at zero
Use 0~5 keys to enter condition at zero, Press the
【M+】 key to confirm:
ZP 1 : 1 division not to display at zero (factory
default)
Z P 2 : 2 division not to display at zero
Z P 3 : 3 division not to display at zero
Z P 4 : 4 division not to display at zero
Z P 5 : 5 division not to display at zero
Press 【M+】key to confirm, press 【CE】key to exit.
Selectable Weighing Unit (Factory Default kg)
Press and hold the【M+】key and press the【UNITS】key until “units 1” for kg appears.
For kg and lb, press the【T】key to move to “units 2”. Press the 【M+】key to confirm
the setup and return to regular weighing mode.
Standard Calibrations
At weighing mode,press and hold the【Z】key,then press the【M+】key,press
【T】key or【Z】key to select ECF-1~ECF-3
ECF-1
Weight
calibration
Press the【M+】key to enter,display internal value, press the
【M+】key to calibration the zero weight after stable,Use 0~9
keys to enter the weight to be calibrated. Place the required weight
mass onto the scale,press【M+】key to confirm the calibration
procedure completed,press the 【CE】key return to previous
setup mode
ECF-2
Zero
calibration
Press the【M+】key to enter,display internal value, press the
【M+】key to calibration the zero weight after stable,press the
【CE】key return to previous setup mode
ECF-3
SPAN
calibration
Press the【M+】key to enter. Use 0~9 keys to enter the weight to
be calibrated. Place the required weight mass onto the scale,
Place the required weight mass onto the scale,press【M+】key
to confirm the calibration procedure completed,press the 【CE】
key return to previous setup mode