00967 Deluxe Force Gauge
Greenlee Textron / Subsidiary of Textron Inc. 8 4455 Boeing Dr. • Rockford, IL 61109-2988 USA • 815/397-7070
Configuration
Under normal circumstances it is not necessary to enter the configuration mode. Directions for this can be found in the
Strain Gage Meter operator’s manual supplied.
To display kilograms instead of pounds change the value for: “RD.S.O., RD2” to 2268
“OT.S.O., RD2” to 3629
In the event any of the default values set by the factory are changed by accident, the Greenlee default values are
listed below.
Menu Topic Setting/Value
INPT (signal input) 100m
DEC.P (decimal point) FFFF.
RD.S.O. (reading scale and offset)
IN1 (no load signal input) ______
RD1 (no load display reading) 0000
IN2 (5000 pound signal input) ______
RD2 (5000 pound display reading) 5000
(2268 for kilograms)
RD.CF (reading, configuration)
R.1 (ratiometric operation) R
R.2 (input resolution) 4
R.3 (filtering) F
R.4 (gross/net peak display) P
S1.CF (setpoint 1 configuration) not active
S2.CF (setpoint 2 configuration) not active
S1DB (setpoint 1 deadband) not active
S1DB (setpoint 2 deadband) not active
OT.CF (output enabled/disabled)
0.1 (output enabled/disabled) E
0.2 (voltage/current output) V
0.3 (net/gross reading) N
OT.S.O. (output scale and offset
RD1 (no load output reading) 0000
OUT.1 (no load output signal) 00.00
RD2 (loaded output reading) 8000
(3629 for kilograms)
OUT.2 (loaded output signal) 08.00
LK.CF (lockout configuration)
SP. (setpoint lockout) E
RS. (reset lockout) E
S1 Dip Switch Positions
1 - Closed
2 - Open
3 - Closed
4 - Open
5 - Open
6 - Closed
7 - Open
8 - Open