16
*1
The voltage display is set to RMS in AC/AC+DC mode and AVG in DC
mode.
*2
AC mode: For an output voltage of 17.5 V to 175 V / 35 V to 350 V and
23 °C ± 5 °C. DC mode: For an output voltage of 25 V to 250 V / 50 V to
500 V and 23 °C ± 5.
*3
An output current in the range of 5 % to 100 % of the maximum current,
and 23 °C ± 5 °C.
*4
An output current in the range of 5 % to 100 % of the maximum peak
current in AC mode, an output current in the range of 5 % to 100 % of
the maximum instantaneous current in DC mode, and 23 °C ± 5 °C. The
accuracy of the peak value is for a waveform of DC or sine wave
*5
For an output voltage of 50 V or greater, an output current in the range
of 10 % to 100 % of the maximum current, DC or an output frequency of
45 Hz to 65 Hz, and 23 °C ± 5 °C.
*6
The apparent and reactive powers are not displayed in the DC mode.
*7
The reactive power is for the load with the power factor 0.5 or lower.
*8
An output voltage in the range of 17.5 V to 175 V / 35 V to 350 V and 23
°C ± 5 °C.
Note
Product specifications are subject to change without
notice.
The spec aforementioned applies to when slew rate
mode is the Time mode.
General Specifications
Interface
Standard
USB
LAN
MAC Address, DNS IP Address, User
Password, Gateway IP Address,
Instrument IP Address, Subnet Mask
EXT
Control
External Signal Input
External Control I/O
Type A: Host, Type B: Device,
Speed: 1.1/2.0, USB-CDC