EasyManua.ls Logo

GW Instek ASR-2000 Series - Page 19

GW Instek ASR-2000 Series
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
Nondestructive
maximum input voltage
±10 V
Gain resolution
0.1 times
Accuracy
±5 % (DC, or 45Hz ~ 65 Hz, gain is at initial value,
with rate voltage output, no load)
EXT: Output voltage (V) = External signal input (V) x Gain (V/V)
Voltage Setting Signal Input (AC-VCA Mode)
Specification
Factory Default
Gain setting range
100 V range: 0.0 to 250.0 times
100
200 V range: 0.0 to 500.0 times
200
Input terminal
BNC connector
Input impedance
1 MΩ
Input voltage range
DC 0 ~ 2.5 V
Nondestructive
maximum input voltage
±10 V
Accuracy
±5 %
External Signal Input (AC+DC-ADD, AC-ADD Mode)
Specification
Factory Default
Gain setting range
100 V range: 0.0 to 250.0 times
100
200 V range: 0.0 to 500.0 times
200
Input terminal
BNC connector
Input impedance
1 MΩ
Input voltage range
±2.5 V (A/D resolution 12 bit)
Nondestructive
maximum input voltage
±10 V
Input frequency range
DC to 999.9 Hz (sine wave)
DC to 100 Hz (other than sine wave)

Other manuals for GW Instek ASR-2000 Series