GOM-804 & GOM-805 Firmware Update Manual
16
5. Select the Program tab. This will allow you to
program the flash on the Stellaris microcontroller.
6. In the Select .bin file area select the slave firmware
.bin file. GOM805_SLAVE_*.BIN
Notice:”Program Address Offset” must be set
2800.
Warning
Failing to set the Program Address Offset to 2800 will
lead to serious consequences in that the unit may fail to
turn on properly and will need to be returned to GW
Instek for repair.
7. Click the Program button to begin the flashing the
microcontroller.
8. When the firmware has been flashed, power off
the instrument & disconnect the USB cable from
the rear panel of the instrument.
9. Power on the instrument. The instrument should
now have the new firmware installed.