GOM-804 & GOM-805 Firmware Update Manual
4
INTRODUCTION
The GOM-804/805 DC Milliohm Meter has upgradable firmware to ensure
your purchase has a long and useful life with updates and improvements.
The meter has two microcontrollers that require firmware updates, a master
microcontroller and a slave microcontroller. The master microcontroller is
used to control the display, the panel keys and SCPI command parsing. The
slave controls the relays and ADC operation for the measurement circuits.
The firmware for each microcontroller must be updated separately and as
such this manual has a master microcontroller update section and a slave
microcontroller update section.
The software used to update each microcontroller is the same
(LMFlashProgrammer.msi), however the interface used and the procedure to
update the controllers is slightly different.
Requirements
Hardware
GDM-804 / GDM-805
USB cable Type A/B
Software
Windows Operating System (XP, Windows 7, 8,
8.1)
Program kit using LM Flash Programmer
DFU driver (device firmware upgrade)
GOM-804 / 805 Firmware
MASTER: GOM805_MASTER_*.BIN
SLAVE: GOM805_SLAVE_*.BIN
Note
Contact your local distributor for latest firmware.