OPERATION
45
3. Press the HI/LO soft-key again or
press the DOWN arrow key to
switch between HI SET and LO
SET.
I / LH O
OR
I R G B A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
1 . 0 0 m AIH ES = 0T
m A
4. Use the scroll wheel to set the HI
SET/LO SET limit*.
ACW (LO)
DCW (LO)
IR (HI)
GB (LO)
0.000mA~041.9mA (GPT-98XX)
0.000mA~109.9mA (GPT-99XX/
99XXA)
0.000mA~010.9mA (GPT-98XX)
0.000mA~020.9mA (GPT-99XX/
99XXA)
0001M~9999M, (GPT-98XX)
0.001G~50.00GΩ, (GPT-99XX/
99XXA)
000.0m~ 649.9m
Note
*Please note that the resolution of the measured
value depends on the resolution of HI SET setting.
Note
The LO SET setting is limited by the HI SET
setting. The LO SET limit cannot be greater than
the HI SET limit.
When setting the current, be aware that a
maximum of 200VA can be set for ACW and 50W
for DCW (GPT-98XX) or 500VA and 100W,
respectively for GPT-99XX/99XXA.