Page #116 background imageLoading...
Page #116 background image
GPT-9000/9000A Series User Manual
116
Display
I R G B A CW D CW
F R E Q = 0 H z 6
0
100
k V
TMR
MT I E = 0 0 3 . 2 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
2M A UN 0 0_2M A N U = * * * - 0 0
m A
Return to Panel Control
Background
When the instrument is remotely controlled all
panel keys except the STOP button are
disabled.
Steps
1. When RMT is on the display, press
the STOP button. The panel goes to
the READY status.
STOP
2. From the READY status the tester
can go into one of two states: TEST
or VIEW.
To put the tester into VIEW status,
press the MANU/AUTO key.
To put the tester in TEST status,
press the START button. This will
start the manual test/automatic
test. For more details on running a
manual test or automatic test, see
pages 64 and 90, respectively.
MANU/AUTO
START
Note
To put the tester back to RMT, simply issue another
remote control command.
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9904

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9904 and is the answer not in the manual?

GW Instek GPT-9904 Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9904
CategoryTest Equipment
LanguageEnglish

Related product manuals