GPT-9000/9000A Series User Manual
28
5. Ensure the earth ground of the line voltage is
properly grounded.
6. Only connect the test leads to the HIGH
VOLTAGE/SOURCE H/SENSE H terminals
before the start of a test. Keep the test leads
disconnected at all other times.
7. Always press the STOP button when pausing
testing.
8. Do not leave the safety tester unattended.
Always turn the power off when leaving the
testing area.
9. When remotely controlling the safety tester,
ensure adequate safety measures are in place to
prevent:
Inadvertent output of the test voltage.
Accidental contact with the instrument during
testing. Ensure that the instrument and DUT are
fully isolated when the instrument is remotely
controlled.
10. Ensure an adequate discharge time for the
DUT.
When DCW or IR tests are performed, the DUT,
test leads and probes become highly charged.
The GPT-9000/9000A has discharge circuitry to
discharge the DUT after each test. The time
required for a DUT to discharge depends on the
DUT and test voltage.
Never disconnect the safety tester before a
discharge is completed.
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9904

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9904 and is the answer not in the manual?

GW Instek GPT-9904 Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9904
CategoryTest Equipment
LanguageEnglish

Related product manuals