OPERATION
75
I<SET
If SOURCE H/L terminals are open or poorly
connected, then an I<SET error will appear on the
screen. Stop the test and re-check the connection
again and try again.
Z E R 0 I R G B A CW D CW
E
F R E Q = 0 H z 6
IAF
G B V = 5 . 6 2 2 V MT I E = 0 0 3 . 2 SR
. 0 m ER F =# 0 00
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
L
I<SET error message
FAIL status
< I S E T
R = 0
Stop the test and perform the zero check again.
Z E R 0 I R G B A CW D CW
E
F R E Q = 0 H z 6
IAF
G B V = 5 . 6 2 2 V MT I E = 0 0 3 . 2 SR
. 0 m ER F =# 0 00
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
L
R = 0 error message
FAIL status
= R 0