LCR-800 Series User Manual
54
! Press F1/F2 key to select the white inverted characters “C_Nom.Val”.
! Key-in Nominal Value through panel keyboard.
! Press
key. Please refer to Figure 8-16 below.
Figure 8-16
Max setting:
! Press F1/F2 key to select the white inverted characters “Max”.
! Key-in SORT Max value through panel keyboard.
! Press
key. Please refer to Figure 8-17 below.
After the value is input, it will proceed conversion of +% automatically.
Figure 8-17
+” setting:
! Press F1/F2 key to select the white inverted characters “+”.
! Key-in the largest percentage of SORT Nominal Value through the panel
keyboard.
! Press
key. Please refer to Figure 8-18 below.
EXIT
BIN1
SORT BY C TOt_Bin:13
:15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.000
D_Min:.0010
CLR
Press
F1.F2
key to the
C_Nom.Val
NEXT
BIN
C_Nom.Val
Press F3 key
can delete the
values input
in error
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val:15.000pF
:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
Press
F1.F2
key to the
Max
NEXT
BIN
Max
Press F3 key
can delete the
values input
in error