LCR-800 Series User Manual
55
After the value is input, it will proceed conversion of Max automatically.
Figure 8-18
Min setting:
! Press F1/F2 key to select the white inverted characters “Min.
! Key-in SORT Min value through panel keyboard.
! Press
key. Please refer to Figure 8-19 below.
After the value is input, it will proceed conversion of -% automatically.
Figure 8-19
-” setting:
! Press F1/F2 key to select the white inverted characters “-”.
! Key-in the smallest minus percentage of SORT Nominal Value through the
panel keyboard.
! Press
key. Please refer to Figure 8-20 below.
After the value is input, it will proceed conversion of Min automatically.
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val:15.000pF
Max:16.500pF+:10.0%
:13.500pF-:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
Press
F1.F2
key to the Max
NEXT
BIN
Min
Press
F3
key
to delete the
values input
in error
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF :10.0%
Min:13.500pF -:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
Press
F1.F2
key to the
+
NEXT
BIN
+
Press
F3
key
to delete the
values input
in error

Related product manuals