Chlorine, Free and Total, LR, continued
4.Press ZERO/SCROLL.
The display will show
“- - - -” then “0.00”.
Remove the blank.
5.Fill a DPD Free Chlorine
Reagent AccuVac®
Ampule or a DPD Total
Chlorine Reagent AccuVac
Ampule with sample.
Note: Keep the tip
immersed until the ampule
fills completely.
6.Quickly invert the
ampule several times to
mix.
Note: A pink color will form
if chlorine is present.
Note: Accuracy is not
affected by undissolved
powder.
5957088.book Page 26 Friday, March 22, 2013 3:54 PM