2—27
User-Entered Calibration, continued
Maximum/Minimum Displayed Value
In meters with absorbance (Abs) ranges, the maximum displayed value and
minimum displayed value is related to the value of the standards entered in a
user calibration.
Measurements that exceed the minimum or maximum standards entered in the
user calibration will return a flashing number indicating “underrange” or
“overrange”. See Error Codes (page 2—12) for more information.
Example 1
For a calibration with the following standards:
S0=0.000
S1=1.000
Maximum Displayed Value 1.000
Minimum Displayed Value 0.000