4.2 Preparation for use
Install the analyzer bottles and stir bar. Refer to the operations manual for the startup procedure.
Section 5 Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
D A N G E R
Electrocution hazard. Remove power from the instrument before doing maintenance or
service activities.
W A R N I N G
Biohazard exposure. Obey safe handling protocols during contact with sample bottles
and sampler components.
W A R N I N G
Multiple hazards. The technician must make sure that the equipment operates safely and
correctly after maintenance procedures.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or
repaired, contact the manufacturer.
5.1 Clean the instrument
C A U T I O N
Fire hazard. Do not use flammable agents to clean the instrument.
N O T I C E
Do not clean the controller compartment heater with liquids of any kind.
If water is not sufficient to clean the controller and the pump, disconnect the controller and move the
controller away from the sampler. Allow sufficient time for the controller and pump to dry before the
parts are re-installed and put back into service.
Clean the sampler as follows:
Refrigerator—clean the condenser fins and coils as needed with a brush or vacuum.
Sampler cabinet and tray—clean the internal and external surfaces of the sampler cabinet with a
damp cloth and mild detergent. Do not use abrasive cleaners or solvents.
20
English

Other manuals for Hach AS950

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach AS950 and is the answer not in the manual?

Hach AS950 Specifications

General IconGeneral
Sample FrequencyProgrammable
TypePortable
Sampling MethodVacuum Pump
Temperature ControlRefrigerated
Bottle Configuration24 x 1L bottles
Power Supply12V DC
CommunicationRS-232
Sampler TypePortable or Stationary
Vertical Lift8.2 m
Enclosure RatingIP67
Number of SamplesUp to 999
Storage Capacity24 samples
Bottle Options1L

Related product manuals