To replace the power supply for the refrigerated sampler, refer to Figure 16.
Figure 16 Power supply replacement
5.6 Disposal
D A N G E R
Child entrapment hazard. Remove the doors on the refrigerated cabinet before disposal.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Section 6 Troubleshooting
6.1 General troubleshooting
Table 2 shows causes and corrective actions for several common problems.
Table 2 Troubleshooting table
Problem Possible cause Solution
No instrument power Problem with the main power
source.
Make sure that AC power gets to
the electrical outlet.
Defective power supply
(refrigerated sampler only)
Replace the power supply.
Defective controller Contact technical support.
26 English

Other manuals for Hach AS950

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach AS950 and is the answer not in the manual?

Hach AS950 Specifications

General IconGeneral
Sample FrequencyProgrammable
TypePortable
Sampling MethodVacuum Pump
Temperature ControlRefrigerated
Bottle Configuration24 x 1L bottles
Power Supply12V DC
CommunicationRS-232
Sampler TypePortable or Stationary
Vertical Lift8.2 m
Enclosure RatingIP67
Number of SamplesUp to 999
Storage Capacity24 samples
Bottle Options1L

Related product manuals