Error Description Solution
E-7 Standard value
error
The standard solution concentration is equal to
another standard solution concentration that is
already entered in the user-entered calibration
curve. Enter the correct standard concentration.
E-9 Flash error The instrument is not able to save data.
Reading
flashes
The reading is
more or less than
the instrument
range.
2
If the reading is less than the instrument range,
make sure that the instrument cap is fully
installed over the cell holder. Measure a blank.
If the blank reading is not zero, set the
instrument to zero again.
If the reading is more than the instrument
range, identify if there is a light blockage in the
cell holder. Dilute the sample. Do the test again.
For factory-calibrated programs, the maximum
and minimum values always equal the factory-
calibrated values and cannot be changed.
1
When an E-1 or E-2 error occurs on a measurement, the display shows
“_.__”. The decimal place depends on the chemistry. If the E-1 or E-2 error
occurs while the instrument is set to zero, set the instrument to zero again.
2
The flashing value will be 10% over the upper test range limit.
Replacement parts
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment malfunction.
The replacement parts in this section are approved by the
manufacturer.