DOC023.98.93068
Polymetron 9526 Conductivity
Certification System
05/2015, Edition 4
Basic User Manual
Basishandbuch
Manuale dell'utente di base
Manuel d'utilisation de base
Manual básico del usuario
Manual básico do utilizador
基本用户手册
Basisgebruikershandleiding
Podstawowa instrukcja obsługi
Peruskäyttöohje
Начальное руководство пользователя
Temel Kullanıcı Kılavuzu

Table of Contents

Questions and Answers

  • R
    Ryan WilkersonAug 19, 2025
    Why does my Hach Analytical Instrument display 'NOT CALIBRATED'?
    • P
      Paul BlackAug 19, 2025
      The Hach Analytical Instrument displays 'NOT CALIBRATED' when the sensor requires calibration. Calibrate the sensor to resolve this issue.
  • C
    Charles TorresAug 21, 2025
    What to do if my Hach Analytical Instrument shows 'CAL OVERDUE'?
    • E
      Evan JenningsAug 21, 2025
      If your Hach Analytical Instrument displays 'CAL OVERDUE', it means the sensor calibration is overdue. You should calibrate the sensor.
  • H
    Hannah RichmondAug 23, 2025
    How to resolve 'MEAS TOO HIGH' error on Hach Polymetron 9526?
    • T
      Tammy MooreAug 23, 2025
      To resolve the 'MEAS TOO HIGH' error on your Hach Analytical Instruments, ensure that the display format is set for the correct measurement range.
  • S
    Sharon BrownAug 26, 2025
    How to fix 'MEAS TOO LOW' error on Hach Polymetron 9526?
    • T
      tuckeraliciaAug 26, 2025
      To fix the 'MEAS TOO LOW' error on your Hach Analytical Instruments, make sure that the sensor is configured for the correct cell constant.
  • K
    Krista BrownAug 28, 2025
    What to do if Hach Analytical Instrument shows 'SENS OUT RANGE'?
    • R
      riggsbrendaAug 28, 2025
      If your Hach Analytical Instrument shows 'SENS OUT RANGE', ensure that the display format is set for the correct measurement range.
  • J
    Jeffrey NolanAug 30, 2025
    How to troubleshoot 'ADC FAILURE' on Hach Polymetron 9526?
    • K
      Keith ObrienAug 31, 2025
      To troubleshoot 'ADC FAILURE' on Hach Analytical Instruments: * Make sure that the sensor module is fully inserted into the controller connector. * Replace the sensor module.
  • B
    Brittany HoustonSep 1, 2025
    Why is my Hach Analytical Instrument showing 'ZERO TOO HIGH'?
    • D
      Desiree MooreSep 2, 2025
      The Hach Analytical Instrument may show 'ZERO TOO HIGH' if the sensor is not held in air during zero calibration or is near radio frequency or electromagnetic interference. Also, ensure that the cable is shielded by metal conduit.
  • K
    Kelly HuberSep 3, 2025
    What to do if Hach Polymetron 9526 Analytical Instruments display 'TEMP TOO HIGH'?
    • K
      kaitlintownsendSep 4, 2025
      If Hach Analytical Instruments display 'TEMP TOO HIGH', make sure that the sensor is configured for the correct temperature element.
  • P
    Patricia SloanSep 6, 2025
    What should I do if my Hach Analytical Instrument displays 'CAL IN PROGRESS'?
    • K
      Kimberly BoydSep 6, 2025
      If your Hach Analytical Instrument displays 'CAL IN PROGRESS', return to calibration.
  • V
    Vickie TravisSep 8, 2025
    What does 'WRONG LINEAR TC' mean on my Hach Analytical Instruments?
    • A
      Angela WillisSep 9, 2025
      The 'WRONG LINEAR TC' error on your Hach Analytical Instruments means that the value must be between 0 and 4%/°C; 0 to 200 °C.

Hach Polymetron 9526 Specifications

General IconGeneral
TypeSilica Analyzer
Power Supply100 - 240 VAC, 50/60 Hz
ParameterSilica
Measurement Range (pH)Not Applicable
Resolution (pH)Not Applicable
CalibrationAutomatic
Measurement MethodColorimetric
Outputs4-20 mA, RS-232
Response TimeTypically 10 minutes