Table 4 Warning list for conductivity sensors (continued)
Warning Description Resolution
WRONG LINEAR TC The user-defined linear temperature
compensation is out of range
The value must be between 0 and
4%/°C; 0 to 200 °C.
WRONG TC TABLE The user-defined temperature
compensation table is out of range
The temperature is above or below
the temperature range defined by
the table.
WRNG USER CONC TABLE The concentration measurement is
outside of the range of the user
table
Make sure that the user table is set
for the correct measurement range.
WRNG BLT-IN TEMP TABLE The measured temperature is
outside of the range of the built-in
temperature compensation table
Make sure that the temperature
compensation is configured
correctly.
WRNG BLT-IN CONC TABLE The concentration measurement is
outside of the range of the built-in
concentration table
Make sure that the concentration
measurement is configured for the
correct chemical and range.
Replacement parts and accessories
Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.
Note: Product and article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts and accessories
Description Item no.
Kit of 3 protection caps for instrument front panel connectors 09126=A=8010
Kit of 2 adapters D6/8 to DN4/6 09126=A=8020
Kit of 2 black protection caps for sample input and output connectors 09126=A=8030
Power supply connector 350=500=004
Tool for disconnecting in/out sampling tubes 578=507=602
DN8 semi-rigid PTFE tubing (per meter) 590=060=080
DN8 semi-rigid PE tubing (per meter) 151400,22387
Cable for the 4-20 mA output (5 meters) 08319=A=0005
Cable for the 4-20 mA output (10 meters) 08319=A=0010
Cable for the 4-20 mA output (20 meters) 08319=A=0020
¾’’ NPT flow chamber in PP with fittings 09126=A=0100
Pt100 temperature simulator (0.1°C accuracy) 037=000=001
Cable for connecting the Pt100 temperature simulator 09125=A=8020
Yearly re-calibration in our factory 09526=A=1000