Table of contents
Specifications on page 3 Startup on page 13
General information on page 5 Maintenance on page 18
Installation on page 9 Troubleshooting on page 19
User interface and navigation on page 13
Additional information
Additional information is available on the manufacturer's website.
Specifications
Specifications are subject to change without notice.
Analyzer
Specification Details
Dimensions Height: 450 mm; Width: 250 mm; Depth: 460 mm
Weight 7 kg (15.4 lb)
Casing protection IP 65 / NEMA4X
Power supply
Standard version: 100-240 VAC 50/60 Hz
Low voltage version: 13-30 VAC 50/60 Hz, 18-42 VDC
Consumption: 25 VA
Measurement category: I (overvoltage less than 1,500 V)
Sample flow rate 20 liters/hour minimum
Sample tubing
Sample inlet and outlet: Diameter 8 mm (or 5/16'') semi-rigid tubing. We recommend the
use of PE tubing if sample temperature is inferior to 70 °C, and PTFE if superior to 70 °C
Connections
Power supply: Use the connector provided in the drawer
Analog output: Use the recommended POLYMETRON cable
Ambient temperature -20 to 60 °C (-4 to 140 °F)
Maximum temperature 100 °C (at atmospheric pressure)
Maximum pressure 10 bar at ambient temperature
Relative humidity 10—90%
Precision
Conductivity: ± 2% of the displayed value
Temperature: ± 0.2 °C
Measurement range
Conductivity: 0.01 μS/cm to 200 μS/cm
Resistivity: 100 MΩ.cm to 5 kΩ.cm
Temperature: -20 to 200 °C (-4 to 392 °F)
Display resolution 0.001 μS/cm or 0.1 MΩ.cm