Questions and Answers

  • A
    Anita HudsonAug 12, 2025
    What to do if the signal is not stable on Hach Analytical Instruments?
    • Q
      qtaylorAug 13, 2025
      If the signal from your Hach Analytical Instruments is not stable, wait for stability to be reached and check the N2 gas flow rate. Also, check for the presence of liquid on the LDO membrane. If liquid is found in the flow meter, dry it by injecting 20 mL of EtOH followed by 5 minutes of N2 gas.
  • A
    andrewoconnorAug 16, 2025
    Why is the response time too long on Hach Lange ORBISPHERE 3100 Analytical Instruments?
    • J
      Jordan SmithAug 16, 2025
      Several factors can cause a long response time. Try the following: * Use the tubes supplied with the instrument, as they are made of low O2 permeable material. * Check the flow rate and connections to ensure they are correct. * If sampling a package, purge the piercer with N2 gas before sampling. * Perform the monthly cleaning procedure.
  • M
    mark35Aug 20, 2025
    What to do if I see measurement spiking on Hach Analytical Instruments?
    • M
      Michael HenryAug 20, 2025
      If you observe measurement spiking, ensure all tubing is tightly and correctly connected to the sample inlet and outlet connections. Also, remove any residual air bubbles from the sample line by turning the sample inlet/outlet valve to the PURGE position for 5 seconds.
  • M
    melissa30Aug 23, 2025
    What to do if the signal is not in range on Hach Analytical Instruments?
    • K
      Kathryn BuchananAug 23, 2025
      If the signal is not in range and the calibration value is very different from the last calibration value, completely open the sample flow adjustment valve. Also, the calibration gas may have too much oxygen; use a quality 5.0 N2 or CO2 bottle. Check for oxygen leaks between the reference bottle and the instrument by increasing the gas flow rate and checking for a decrease in partial pressure (mbar).
  • D
    Dawn ReedAug 27, 2025
    How to resolve degassing in the flow meter of Hach Lange ORBISPHERE 3100 Analytical Instruments?
    • A
      Andrew SummersAug 28, 2025
      To resolve degassing in the flow meter, adjust the flow rate to between 150 and 200 mL/min. Also, verify that the filter (if you are using one) is not blocked.
  • M
    martinwesleyAug 31, 2025
    What to do if silver bead does not move freely in the sample on Hach Lange ORBISPHERE 3100 Analytical Instruments?
    • D
      douglasjessicaAug 31, 2025
      If the silver bead does not move freely in the sample, perform the monthly cleaning procedure.
  • R
    Robert DavisSep 4, 2025
    How to resolve channel out error on Hach Lange ORBISPHERE 3100 Analytical Instruments?
    • C
      Cassidy WalterSep 4, 2025
      To resolve a channel out error, change the LDO spot. In other cases, contact Hach Lange service.
  • B
    Brandon GarciaSep 8, 2025
    What to do if the display appears frozen on Hach Lange ORBISPHERE 3100 Analytical Instruments?
    • K
      Katherine HarrellSep 8, 2025
      If the display appears frozen with no active cursor, check the measurement mode to see if it is set to continuous or sample.

Hach Lange ORBISPHERE 3100 Specifications

General IconGeneral
Temperature Range-5 °C to +50 °C (23 °F to 122 °F)
Pressure Range0 to 10 bar absolute
Communication InterfaceRS232 or RS485
ProtectionIP65
Output4-20 mA
Measurement ParametersDissolved Oxygen, Ozone, Hydrogen Peroxide