A bit value of 1 shows the error or status condition that is true. For example, if Bit 0 has the value of
1, an error has occurred during the last calibration.
Table 2 Error messages
Bit Message Indication
0 Measurement calibration error An error has occurred during the last calibration
1 Electronic adjustment error An error has occurred during the last electronic calibration
2 Cleaning error The last cleaning cycle failed
3 Measuring module error A failure has been detected in the Measurement Module
4 System re-initialization error Some settings are inconsistent and have been reset to factory
defaults
5 Hardware error A general hardware error has been detected
6 Internal communication error A communication failure within the device has been detected
7 Humidity error Excessive humidity has been detected within the device
8 Temperature error Temperature within the device exceeds a specified limit
9 — —
10 Sample warning Some action is required with the sample system
11 Questionable calibration warning The last calibration may not be accurate
12 Questionable measurement warning One or more of the device measurements are out of range or are of
questionable accuracy
13 Safety warning A condition has been detected which may result in a safety hazard
14 Reagent warning The reagent system requires attention
15 Maintenance required warning The device requires maintenance
Table 3 Status indicator messages
Bit Message Indication
0 Calibration in progress The device is in a calibration mode. Measurements may not be valid.
1 Cleaning in progress The device is in a cleaning mode. Measurements may not be valid.
2 Service/Maintenance menu The device is in a service or maintenance mode. Measurements may not be
valid.
3 Common error The device has recognized an error. See Error Register for Error Class.
4 Measurement 0 Quality Bad Precision of measurement is out of specified limits.
5 Measurement 0 Low Limit Measurement is below the specified range.
6 Measurement 0 High Limit Measurement is above the specified range.
7 Measurement 1 Quality Bad Precision of measurement is out of the specified limits.
8 Measurement 1 Low Limit Measurement is below the specified range.
9 Measurement 1 High Limit Measurement is above the specified range.
10 Measurement 2 Quality Bad Precision of measurement is out of the specified limits.
11 Measurement 2 Low Limit Measurement is below the specified range.
12 Measurement 2 High Limit Measurement is above the specified range.
18 English