Questions and Answers

  • C
    Carrie RuizAug 18, 2025
    Why does my Hach Measuring Instruments show E2 error during calibration?
    • A
      alexandra12Aug 18, 2025
      The E2 error during calibration on your Hach Measuring Instruments indicates that the difference between the constants is greater than 30%. To resolve this, try the following: * Insert the probe into the appropriate standard and read again. * Clean the probe and the membrane. * Ensure there are no air bubbles in the membrane. * Disconnect and then reconnect the probe. * Connect a different probe to check if the issue is with the probe or the meter. * If necessary, connect a new probe.
  • J
    Janice ElliottAug 20, 2025
    Why Hach sensION+ EC5 Measuring Instruments display E3 error during calibration?
    • R
      Rachel MorenoAug 21, 2025
      The E3 error during calibration on your Hach Measuring Instruments indicates that the cell constant is less than 0.05 cm-1. To fix it: * Insert the probe into the appropriate standard and read again. * Clean the probe and the membrane. * Ensure there are no air bubbles in the membrane. * Disconnect and then reconnect the probe. * Connect a different probe to verify if the problem is with the probe or the meter.
  • S
    Sean GravesAug 23, 2025
    What to do if my Hach sensION+ EC5 Measuring Instruments show 'Measurement out of range'?
    • C
      Cynthia FlowersAug 23, 2025
      If your Hach Measuring Instruments display 'Measurement out of range', try these steps: * Insert the probe into the appropriate standard and read again. * Clean the probe and the membrane. * Make sure there are no air bubbles in the membrane. * Disconnect the probe and then connect it again. * Connect a different probe to determine if the problem lies with the probe or the meter.
  • J
    jonesjustinAug 26, 2025
    What to do if my Hach Measuring Instruments displays E4 error?
    • Y
      Yesenia RichAug 26, 2025
      The E4 error on your Hach Measuring Instruments means the buffer solution is not recognized. Here's how to troubleshoot: * Clean the probe and the membrane. * Ensure there are no air bubbles in the membrane. * Disconnect and reconnect the probe. * Try a different probe to see if the problem is with the probe or the meter. * Verify that the buffer used matches the buffer specified in the configuration. * Replace the buffer solution. * Check the temperature specification in the configuration.
  • P
    patricklauraAug 27, 2025
    How to fix E2 error on Hach sensION+ EC5 when probe current exceeds limit?
    • J
      jessejacksonAug 27, 2025
      If your Hach Measuring Instruments shows an E2 error because the probe current exceeded the upper limit during measurement (250 nA), try the following: * Clean the probe and the membrane. * Make sure there are no air bubbles in the membrane. * Disconnect the probe and then connect it again. * Connect a different probe to see if the issue is with the probe or the meter.
  • M
    Michael FletcherAug 29, 2025
    What does E5 error mean on Hach sensION+ EC5 Measuring Instruments?
    • A
      adamhughesAug 30, 2025
      The E5 error on your Hach Measuring Instruments indicates that the same buffer solutions are being used. To resolve this: * Clean the probe and the membrane. * Make sure there are no air bubbles in the membrane. * Disconnect the probe and then connect it again. * Connect a different probe to check if the issue is with the probe or the meter. * Ensure that the buffer used matches the buffer specified in the configuration. * Replace the buffer solution. * Verify the temperature specification in the configuration.
  • J
    john41Sep 1, 2025
    How to fix unstable reading on Hach sensION+ EC5?
    • J
      Jennifer BairdSep 1, 2025
      If you're experiencing an unstable reading during a stability measurement or calibration with your Hach Measuring Instruments, make sure that the probe is properly immersed in the sample.
  • J
    Jennifer CobbSep 3, 2025
    How to fix E3 error (slope out of range) on Hach sensION+ EC5 Measuring Instruments?
    • C
      colleenjimenezSep 4, 2025
      If your Hach Measuring Instruments displays an E3 error indicating that the slope is out of range (accepted values ± 58 mV), connect a new probe.
  • K
    Kelly ClarkSep 5, 2025
    What does it mean when my Hach Measuring Instruments show E2 error sensitivity out of range?
    • L
      Leah EllisSep 5, 2025
      If your Hach Measuring Instruments shows an E2 error because the pH probe sensitivity is out of range (accepted values 70–105%), make sure that the correct probe is connected to the meter. If the issue persists, connect a new probe.
  • K
    Kathryn TaylorSep 8, 2025
    How to resolve E6 error on Hach sensION+ EC5 Measuring Instruments?
    • J
      Jill NelsonSep 8, 2025
      The E6 error on your Hach Measuring Instruments indicates that the calibration solutions have different temperatures. Ensure that the calibration solutions have the same temperature.

Summary

Safety Information

Understanding Hazard Symbols

Explains hazard symbols and their associated risks.

Installation

Install the Batteries

Step-by-step guide for inserting batteries into the meter.

Connect a Probe

User Interface and Navigation

Keypad Description

Explains the functions of each button on the meter's keypad.

Display Description

Navigation

Operation

Calibration

Calibration Procedure

Step-by-step guide for calibrating the meter.

Instrument Adjustments and Settings

Maintenance

Replace the Batteries

Battery Replacement

Instructions for safely replacing the meter's batteries.

Troubleshooting

Hach sensION+ EC5 Specifications

General IconGeneral
BrandHach
ModelsensION+ EC5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals