Why does my Hach Measuring Instruments show E2 error during calibration?
- Aalexandra12Aug 18, 2025
The E2 error during calibration on your Hach Measuring Instruments indicates that the difference between the constants is greater than 30%. To resolve this, try the following: * Insert the probe into the appropriate standard and read again. * Clean the probe and the membrane. * Ensure there are no air bubbles in the membrane. * Disconnect and then reconnect the probe. * Connect a different probe to check if the issue is with the probe or the meter. * If necessary, connect a new probe.