Manual
Manual
Manual de instruções
Manuel d’utilisateur
操作手册
07/2010, Edition 1
07/2010, Edición 1
07/2010, Edição 1
07/2010, Edition 1
2010
7月,第一版
DOC022.97.90248

Other manuals for Hach sensION+ pH3

Questions and Answers

  • Q
    qkimJul 31, 2025
    What to do if pH is out of range in Hach Laboratory Equipment?
    • S
      stevenmillerJul 31, 2025
      If the pH is out of range on your Hach Laboratory Equipment, examine the probe. Clean it thoroughly, ensure there are no air bubbles in the membrane, and shake it like a thermometer. You may also want to connect a different probe to determine whether the issue stems from the probe itself or the meter.
  • T
    Tricia VelasquezAug 4, 2025
    What to do if the electrode is in poor condition in Hach Laboratory Equipment?
    • J
      Jacob DavisAug 4, 2025
      If you suspect the electrode in your Hach Laboratory Equipment is in poor condition, examine the probe. Clean the probe, ensuring there are no air bubbles in the membrane. Shake the probe like a thermometer. To determine if the problem lies with the probe or the meter, connect a different probe.
  • K
    Kristopher FloresAug 8, 2025
    What to do if Hach Laboratory Equipment shows unstable reading?
    • M
      Michael HoffmanAug 9, 2025
      If you are experiencing an unstable reading (Time t> 100 s) on your Hach Laboratory Equipment, repeat the measurement. Examine the probe: clean it thoroughly, ensure there are no air bubbles in the membrane, and shake it like a thermometer. Connect a different probe to verify if the problem is with the probe or the meter. Also, ensure that the membrane and the diaphragm are properly immersed in the sample.
  • R
    robertramirezAug 13, 2025
    What to do if Time is more than 150 s in Hach Laboratory Equipment?
    • N
      Nicholas HarrisonAug 13, 2025
      If Time is > 150 s on your Hach Laboratory Equipment, make sure that the membrane and the diaphragm are properly immersed in the sample. Examine the temperature. Examine the probe: clean the probe, ensure there are no air bubbles in the membrane and shake the probe like a thermometer. Try connecting a different probe to verify if the problem is with the probe or meter.
  • Z
    Zachary SimpsonAug 17, 2025
    What to do if buffer or electrode is in poor condition in Hach Laboratory Equipment?
    • D
      dtaylorAug 17, 2025
      If you suspect that the buffer or electrode is in poor condition with your Hach Laboratory Equipment, repeat the measurement. Examine the probe: clean the probe, making sure there are no air bubbles in the membrane. Shake the probe like a thermometer. Connect a different probe to verify if the problem is with the probe or meter. Examine the buffer solution: Make sure that the buffer used matches the buffer specified in configuration; make sure of the temperature specification in configuration; use new buffer solution.
  • J
    Joshua EdwardsAug 21, 2025
    What to do if temperature is out of range in Hach Laboratory Equipment?
    • J
      jonesashleyAug 22, 2025
      If the temperature is out of range on your Hach Laboratory Equipment, examine the temperature sensor. Connect a different probe to verify if the problem is with the probe or the meter.
  • J
    Jonathan SmithAug 27, 2025
    How to fix asymmetry greater than 58 mV on Hach Laboratory Equipment?
    • C
      carolwrightAug 27, 2025
      If the asymmetry is greater than 58 mV on your Hach Laboratory Equipment, it indicates that the slope is out of the accepted range (± 58 mV). Repeat the measurement. Connect a new probe.

Summary

Technical Specifications

General Information

Safety Information

Provides essential safety guidelines, including hazard definitions and precautionary label explanations.

Instrument Description

Brief overview of the sensION™ PH3 instrument and its purpose.

Instrument Composition

Details the components included with the instrument and available accessories.

Installation

Assemblage

Step-by-step guide for assembling the instrument and its components.

Connections

Details how to connect various probes, electrodes, and power supply to the instrument.

Operation

Operation Description and Keys

Overview of the instrument's keys and their functions.

Start-up Procedure

Instructions for the initial setup and subsequent power-on procedures.

General Operation Diagram

Visual representation of the instrument's operational flow and menu access.

pH and mV Calibration

Procedure for calibrating pH and mV measurements using buffer solutions.

pH and mV Measurements

How to perform pH and mV measurements, including stability and continuous modes.

Other Operation Options

Configuration settings for date, time, language, display contrast, and units.

Recognized Buffers

List of recognized buffer solutions and their pH/mV values at different temperatures.

Maintenance

Cleaning the Instrument

Guidelines for cleaning the instrument's housing and accessories.

Sensor Cleaning

Instructions for cleaning the instrument's sensors.

Accessories and Spares

Limited Warranty

Contact Information

Hach sensION+ pH3 Specifications

General IconGeneral
Measurement Range0 to 14 pH
Accuracy±0.01 pH
Resolution0.01 pH
DisplayLCD
IP RatingIP67
Reference SystemAg/AgCl
Shaft MaterialGlass
ConnectorBNC
Temperature CompensationAutomatic
Weight300 g
JunctionPTFE