5. Operation
5.4. Calibrating the redox electrode
Important note: Before turning on the instrument, check that the Redox
electrode is connected.
Redox electrodes do not have a significant drift of the potential, which
is why they are typically not calibrated. However, some deviations of
potential may occur after continued use of the electrode as a result of
changes to the metal surface or contamination of the reference elec-
trode.
5.4.1. Preparation
1. Fill the tube with the 200 mV standard solution (note the fill level).
2. Unscrew the electrode protector containing the electrolyte and rinse
the electrode with deionized water.
3. Screw the electrode on to the tube containing the standard.
5.4.2. Calibration
Stir gently holding the sensor by the handle. Do not hold it by the tube to
avoid heating of the standard solution.
Follow the diagram.
Notes: If the measuring instrument is not used for 5 minutes, it shuts off
automatically.
If errors occur during calibration, an error message will appear on the
display screen (see pg. 19).
During redox calibration, the temperature of the standard must be manu-
ally entered (by pressing
during the process).
ml P/N.
LZW9462
ml
ml
ml
ml
P/N.
LZW94
94
2 3
Screw the
electrode
to the
standard
tube.
1
sensION+ PH1_HACH_0710.indd ANG:12sensION+ PH1_HACH_0710.indd ANG:12 7/9/10 13:09:037/9/10 13:09:03

Table of Contents

Other manuals for Hach sensION+ PH1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sensION+ PH1 and is the answer not in the manual?

Hach sensION+ PH1 Specifications

General IconGeneral
BrandHach
ModelsensION+ PH1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals