EasyManua.ls Logo

Hach TSS sc - Appendix A Modbus Register

Hach TSS sc
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
AppendixA Modbus register
Table 3 Sensor Modbus registers
Tag name Group name Register Data type Length R/W Description
TURBIDITY FNU Measurement 40001 Float 2 R Turbidity in FNU
TURBIDITY NTU Measurement 40001 Float 2 R Turbidity in NTU
TURBIDITY TEF Measurement 40001 Float 2 R Turbidity in TEF
TURBIDITY FTU Measurement 40001 Float 2 R Turbidity in FTU
TURBIDITY EBC Measurement 40003 Float 2 R Turbidity in EBC
SOLID mg/L Measurement 40005 Float 2 R Solid in mg/L
SOLID ppm Measurement 40005 Float 2 R Solid in ppm
SOLID g/L Measurement 40007 Float 2 R Solid in g/L
SOLID % Measurement 40009 Float 2 R Solid in percent
Reserved Reserved 40011 Unsigned integer 1 R Spare
SET PARAMETER Configuration 40012 Unsigned integer 1 R/W Parameter
UnitTM Unit 40013 Unsigned integer 1 R/W Turbidity unit
UnitDS Unit 40014 Unsigned integer 1 R/W Solid unit
OFFSET Calibration 40015 Float 2 R/W Turbidity offset
Factor TRB Calibration 40017 Float 2 R/W Turbidity factor
Factor TS Calibration 40019 Float 2 R/W Solid factor
Reserved Reserved 40021 Unsigned integer 1 R Reserved
RESPONSE TIME Configuration 40022 Unsigned integer 1 R/W Response time
CLEAN. INTERVAL Configuration 40023 Unsigned integer 1 R/W Cleaning interval
LOGGER INTERVAL Configuration 40024 Unsigned integer 1 R/W Logger interval
Outputmodekal Service 40025 Unsigned integer 1 R/W Output mode "Calibrate"
Outputmodesrv Service 40026 Unsigned integer 1 R/W Output mode "Service"
EDITED NAME Configuration 40027 String 8 R/W Measurement location
PROFILE COUNTER Configuration 40035 Unsigned integer 1 R/W Profile counter
SERIAL NUMBER Configuration 40036 String 6 R/W Serial number
CAL. DATE Configuration 40042 Time2 2 R Date of factory calibration
TURBIDITY Calibration 40044 Float 2 R
Turbidity sensor measurement
value
SOLID Calibration 40046 Float 2 R Solid sensor measurement value
PROGRAM Maintenance 40048 Float 2 R Application version
BOOTPROGR. Maintenance 40050 Float 2 R Bootloader version
STRUCTURE Maintenance 40052 Unsigned integer 1 R Structure driver version
FIRMWARE Maintenance 40053 Unsigned integer 1 R Register driver version
CONTENT Maintenance 40054 Unsigned integer 1 R Firmware driver version
FormatMinFNU Configuration 40055 Float 2 R Turbidity lower limit in FNU
FormatMaxFNU Configuration 40057 Float 2 R Turbidity upper limit in FNU
FormatMinEBC Configuration 40059 Float 2 R Turbidity lower limit in EBC
FormatMaxEBC Configuration 40061 Float 2 R Turbidity upper limit in EBC
FormatMinGL Configuration 40063 Float 2 R Solid lower limit in g/L
FormatMaxGL Configuration 40065 Float 2 R Solid upper limit in g/L
FormatMinMGL Configuration 40067 Float 2 R Solid lower limit in mg/L
FormatMaxMGL Configuration 40069 Float 2 R Solid upper limit in mg/L
FormatMinPR Configuration 40071 Float 2 R Solid upper limit in percent

Table of Contents

Related product manuals