20
Operation
4.5 CALIBRATE
Note: Turbidity measurement has been calibrated by the manufacturer — it does not need to be
calibrated again.
Note: It is imperative to calibrate for solid matter measurement (refer to section 4.5.2 Calibration of
the SOLID (TS) parameter).
The zero point for turbidity and solid matter measurement has been set in the sensors by
the manufacturer.
TEST/MAINT
PROBE INFO
SENSOR NAME Device name
EDITED NAME
SERIAL NUMBER
TURBIDITY 0,001 to 9999 FNU
SOLID 0,001 to 500 g/L
MODEL NUMBER Item no. Sensor
CODE VERS Sensor software
PROFILE
PROFILE COUNTER Counter 20,000 backwards
RESET CONFIG MANUAL RESET, security prompt
COUNTERS
MANUAL RESET. PRESS ENTER: security prompt
TEST/MAINT: COUNTER X DAYS BACKWARDS,
GASKET (GASK.): COUNTER X DAYS BACKWARDS,
TOTAL: OPERATING HOURS COUNTER,
MOTOR: WIPE CYCLE COUNTER
INTERVAL Default for maintenance counter
SERVICE
WIPE
SIGNALS Explanation: refer to service manual
S5E1
S5E3
S6E1
S6E3
S5E2
S5E4
S6E2
S6E4
SET OUTMODE Equipment output behavior in the SERVICE menu
HOLD
ACTIVE
SET TRANSFER
SELECTION
Service access
4.4.2 SENSOR setup
SELECT SENSOR (if there is more than one sensor)