1 Active mode & patient group.
2 Main controls. Touch the Controls button (3) to display all controls for the selected mode.
3 Window tabs. Open the associated windows.
4 Input power. Shows available power sources.
5 Alarm silence indicator and countdown. Shows whether alarm silence has been activated,
and displays the remaining silence time.
6 Graphic display. Shows a user-selectable waveform or an Intelligent Panel graphic
(Dynamic Lung, ASV graph, Vent Status).
7 Main monitoring parameters (MMP). View other numeric parameters from the monitored
parameter windows.
8 Message bar. Displays color-coded alarm messages. If an alarm is active, view the alarm
buffer by touching the message bar.
9 Pressure/time waveform. Always displayed.
The red line is the Pmax high pressure alarm setting.
The blue line is the pressure limit automatically 10 cmH2O below the Pmax alarm
setting.
The pink triangles indicate the patient is triggering a breath.
The
Freeze button freezes the graphic for up to 30 s.
10 Alarm indicator (i-icon). Touch the icon to view information about alarms.