INSTRUCTION MANUAL
HI98301
TDS Tester
Warranty
These meters are warranted for a period of one year against defects in
workmanship and materials when used for their intended purpose and
maintained according to instructions. Probes are warranted for a period
of six months. This warranty is limited to repair or replacement free of
charge. Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered. If service is required, contact your local Hanna
Instruments Office. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization (RGA) number from the Technical Service department
and then send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
Solutions
Code Description
HI70032P
1382 ppm (mg/L) calibration solution,
20 mL sachets (25 pcs.)
HI7032M 1382 ppm (mg/L) calibration solution, 230 mL bottle
HI7032L 1382 ppm (mg/L) calibration solution, 500 mL bottle
Accessories
IST98301N 09/18
To obtain the highest accuracy for measurements it is important to follow
these tips:
Calibration is only as good as the solution being used. The calibration
solution values change over time. Fresh solution should be used for
each calibration.
The probe should be rinsed with purified water each time before placing
in calibration solution or sample to be tested.
Care and Maintenance
Error messages
During user calibration, if the reading is out of the accepted range, the
meter will display “---- Err.i.e. 2000 ppm blinking means the limit of the
measured value is out of range. If the measured temperature is higher than
50.0 ºC or lower than 0.0 ºC, the 50.0 ºC or 0.0 ºC temperature value will blink
on the LCD in measure mode.
2. Place the probe in calibration solution. The probe tip should be centered
in the solution and submersed 1.18” (3 cm). The meter will automatically
recognize the solution.
3. If the solution is not recognized or is out of the accepted range “---- Err“
is displayed.
4. If the calibration solution is recognized “REC“ is displayed until the
reading is stable and the calibration is accepted.
After acceptance, the “Stor“ message is displayed and meter returns to
measurement mode.
Calibration Procedure
Measurement and Calibration
For better accuracy, frequent calibration of the meter is recommended. In
addition, the instrument must be recalibrated:
When high accuracy is required.
At least once a month.
Measurement
Press the ON/OFF button to turn the meter ON. Place the probe to the
desired solution. The conductivity value, automatically compensated for
temperature, will be shown on the LCD.
The measured temperature will be shown on the secondary LCD.
Note: Before taking any measurement make sure the meter has been
calibrated.
Press CAL button to enter calibration mode from measurement mode.
The meter will enter the calibration mode, displaying “ppm 1382 USE”
with CAL tag blinking.
1. Pour 2” (5 cm) of standard solution into a container.
Clear Calibration
Place meter in calibration mode. Press ON/OFF and “CLR” is displayed. The
meter will now be at default calibration.
DiST 1
Battery Replacement
To change the CR2032 Li-ion battery, turn the battery cover located on the
back of the meter counterclockwise to unlock. Remove cover and replace with
new battery + side facing up.
Note: Batteries should only be replaced in a safe area using the
battery type specified in this instruction manual. Old batteries should
be disposed in accordance with local regulations.
The meter features a low battery indicator. When the battery is running low
(under 10%), the battery indicator will blink on the LCD. When the battery
is discharged “dEAd bAtt” will be displayed on the LCD for 2 seconds and
the meter will turn off.
0
0
Before using Hanna Instruments products, make sure that they are entirely
suitable for your specific application and for the environment in which they
are used. Any variation introduced by the user to the supplied equipment
may degrade the instrument’s performance. For yours and the instrument
safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments DiST 1 and is the answer not in the manual?

Hanna Instruments DiST 1 Specifications

General IconGeneral
Measurement Range (Conductivity)0 to 1999 µS/cm
Resolution (TDS)1 ppm
Resolution (Conductivity)1 µS/cm
Resolution (Temperature)0.1 °C
Accuracy (Temperature)±0.5 °C
Measurement Range (TDS)0 to 1999 ppm
Measurement Range (Temperature)0 to 50 °C
Accuracy±2% of reading
Temperature CompensationAutomatic
Probe TypeBuilt-in
Battery Type4 x 1.5V AAA
Environment0 to 50 °C

Summary

Measurement and Calibration

Measurement Operation

Turn on meter, place probe in solution, read conductivity and temperature.

Calibration Procedure

Steps to enter calibration, immerse probe in solution, and await recognition.

Clear Calibration

Procedure to reset meter calibration to default settings.

Error Messages

Explains error messages like "---- Err" for out-of-range values.

Care and Maintenance

Maintenance Tips

Tips for accuracy: use fresh solutions and rinse probe between uses.

Accessories and Battery

Calibration Solutions

Lists available calibration solutions with codes and descriptions.

Battery Replacement

Instructions for replacing the CR2032 battery and safety notes.

User Recommendations and Warranty

User Recommendations

Advice on suitability for application and avoiding hazardous environments.

Warranty Information

Details the one-year warranty for defects and exclusions.

Preliminary Examination and Preparation

Preliminary Examination

Instructions to examine the meter for damage upon receipt and check supplied items.

Preparation Steps

Steps to prepare the meter for use, including removing the cap and immersing the probe.

Operation and Specifications

General Operation

Explains how to turn the meter on and use it for measurements.

Technical Specifications

Lists technical specifications like range, resolution, accuracy, and battery life.

Operational Guide

Turn On and Battery Status

Details how to turn the meter on and check battery status.

Enter Calibration Mode

Instructions for entering calibration mode by pressing the CAL button.

Enter Setup Mode

Describes how to access setup mode by removing the battery cover and pressing the Setup button.

Meter Setup Options

How to change measurement units (°C/°F) and auto-off timer.

Return to Measurement Mode

Instructions to return to normal measurement mode after setup or calibration.