Range*
‑2.00 to 16.00 pH
±825 mV (pH‑mV)
0.00 to 6.00 mS/cm (EC) **
0 to 3000 ppm (0.5 CF)/0 to 3990 ppm (0.7 CF)
‑5.0 to 105.0 °C / 23.0 to 221.0 °F
Resolution
0.01 pH
1 mV (pH‑mV)
0.01 mS/cm
10 ppm (mg/L)
0.1 °C / 0.1 °F
Accuracy
@ 25°C / 77°F
±0.02 pH
±1 mV (pH‑mV)
±2% F.S.
±2% F.S.
±0.5 °C / ±1.0 °F
Temperature
compensation
pH ‑ Automatic
EC ‑ Automatic, with β=1.9%/ °C
pH calibration
Automatic, 1 or 2 point, with: pH 4.01; pH 7.01;
pH 10.01
‑ one‑point calibration using quick calibration
solution
EC calibration
Automatic, one‑point at 1.41 mS/cm or
5.00 mS/cm
‑ one‑point calibration using quick calibration
solution
TDS conversion
factor
0.5 (500 ppm) or 0.7 (700 ppm)
Probe (included)
HI1285‑7 pH/EC/TDS/temperature sensor, DIN
connector and 1 m (3.3’) cable
Battery type / life
1.5V AAA (3 pcs.)
approx. 600 hours of continuous use
Auto‑off
user selectable: after 8 min, 60 min or disabled
Environment
0 to 50 °C (32 to 122 °F) RH max. 100%
Meter dimensions
154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Weight
(with batteries)
196 g (6.91 oz.)
Casing ingress
protection rating
IP67
*the pH range is limited from 0 to 13 pH and the temperature range
from 0 to 50 °C (32 to 122 °F) using HI1285‑7 probe.
**mS/cm is displayed as mS on the display.
***TDS conversion factor: 1000 μS/cm = 500 ppm with 0.5 CF.
SPECIFICATIONS
SPECIFICATIONS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments GroLine HI9814 and is the answer not in the manual?

Hanna Instruments GroLine HI9814 Specifications

General IconGeneral
BrandHanna Instruments
ModelGroLine HI9814
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals