7
Each meter is supplied complete with a 9V battery. Remove the
battery compartment cover on the back of the meter (see
page 16). Install the battery while observing its polarity.
Connect the probe to the meter securely
by aligning the pins with the socket
and pushing the plug in.
Make sure that the meter has been
calibrated before taking any measurements (see “Calibration”
section).
Immerse the conductivity probe into
the sample, with the holes on the
shaft completely submerged.
If possible, use plastic beakers or containers
to minimize any EMC interference.
Tap the probe lightly on the bottom of the beaker to remove any
air bubbles which may be trapped inside the PVC sleeve.
Turn the instrument on by pressing the On/Off key.
For HI 8633 and HI 8734:
Take the temperature of the solution
with a ChecktempC or another accurate
thermometer following LCD indication.
Press and hold down the Temp key to display the temperature
and adjust the temperature knob to that of the solution e.g.
20°C.
For HI 8733:
Adjust the TEMPERATURE COEFFICIENT
knob to 2% to compensate for the
temperature effect of average solutions
(to determine exact value for a particular
solution, see page 15).
Select the appropriate measurement range.
OPERATIONAL GUIDE
www. .com
information@itm.com1.800.561.8187

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 8733 and is the answer not in the manual?

Hanna Instruments HI 8733 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 8733
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals