GLP is a set of functions that allows storage and retrieval of data
regarding the maintenance and status of the probe.
All data regarding DO calibration is stored for the user to review when
necessary.
LAST DO CALIBRATION DATA
The last DO calibration data is stored automatically after a successful
calibration. To view the DO calibration data, press GLP when the
instrument is in DO measurement mode.
The instrument will display the date (mm.dd)
and the time (hh:mm) of the last calibration.
Press Setup to view the year for a few
seconds.
Press the ARROW keys to view the next calibration parameter.
Pressing the key:
The calibration standards.
Press Setup to view the temperature of
the calibration for a few seconds.