26
SINGLE SAMPLE READ
Remove the cuvet from the instrument.
Remove the cap.
Add the content of one packet of the specific test
reagent, for:
Free Chlorine Total Chlorine
1 packet of or 1 packet of
HI 93701-0 HI 93711-0
Replace the cap and shake gently for 20 seconds
(or 2minutes in case of seawater analysis).
Replace the cuvet into the holder and ensure
that the mark on the glass is matched with the
mark on the instrument top. Close the lid.
Hold READ/TIMER
for 3 seconds. The display
will show the hourglass blinking and the
countdown prior to measurement.
Alternatively wait for:
Free Chlorine Total Chlorine
1 minute or 2 minutes and 30 seconds
and then just press READ/TIMER
.
In both cases blinking dashes will appear during
measurement.
The instrument directly displays the concentration
in mg/L of free or total chlorine.
Note:If the value is over range, the maximum
value (5.00 mg/L) will blink.