If the calculated calibration coefficients are outside
a certain range the “CAL Err” message is displayed.
CALIBRATION DELETION
HI 93414 is factory calibrated. It is possible to
restore factory calibration by deleting the last
performed calibration.
To delete last calibration, follow the next steps:
Enter the GLP feature by pressing RCL/GLP for
three seconds.
The date of the last calibration will be displayed
on the LCD.
Press READ/TIMER
to see the information
related to calibration. The last panel is the one
with delete calibration.
Press SETUP/DEL to delete the current calibration.
The instrument will display “del done” for a second
and the calibration is deleted, then the instrument
will automatically return to measurement mode.