EasyManuals Logo
Home>Hanna Instruments>Measuring Instruments>HI 96728C

Hanna Instruments HI 96728C User Manual

Hanna Instruments HI 96728C
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
13
Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement. Remove the cuvette.
Add one packet of the HI 93728-0
reagent.
Replace the cap and immediately shake
vigorously for exactly 10 seconds by
moving the cuvette up and down. Continue
to mix by inverting the cuvette gently
and slowly for 50 seconds, while taking
care not to induce air bubbles. A deposit
could remain but it dose not affect the
measurement. Time and way of shaking
could sensitively affect the measurement.
Replace the cuvette into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/
/TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
Distributed by Applied Membranes, Inc.
www.appliedmembranes.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 96728C and is the answer not in the manual?

Hanna Instruments HI 96728C Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 96728C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals