Alternatively, wait for 3 minutes then just
press READ/ /TIMER. In all cases, the
lamp, cuvette and detector icons will
appear on the display, depending on the
measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
nitrate-nitrogen on the LCD.
To convert the reading to mg/L of nitrate
(NO
3
-
), multiply by the factor 4.43.
INTERFERENCES:
Interference may be caused by:
Ammonia and amines, as urea and primary aliphatic amines. Chloride above 100 mg/L (negative
interference). Chlorine above 2 mg/L (positive interference). Copper (it must be absent). Iron
(III) (positive interference). Strong oxidizing and reducing substances. Sulfide (it must be
absent).
Note: To ensure accurate results, preform the tests at room temperature, between 18 °C and 28
°C (65 °F to 83 °F).
For best result: Intensely colored samples or suspended matter in large amounts will cause
interference. Therefore sample should be adequately treated before performing the test.
Distributed by Applied Membranes, Inc.
www.appliedmembranes.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 96728C and is the answer not in the manual?

Hanna Instruments HI 96728C Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 96728C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals