EasyManuals Logo

Hanna Instruments HI2550 User Manual

Hanna Instruments HI2550
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
50
GLDxxxff: Requests the records of the “xxx” lot number. The records are
sent in frames of 10 records; “ff” is the frame number (01
first frame). (Example: Lot 13 has 53 records. The records will
be sent in 6 frame, 5 with10 records and 1 with 3 records.)
Command Parameters:
xxx Lot number (eq: for lot number 1 xxx = 001)
ff requested frame number ‑ first frame is labeled 01
Errors: “Err3” log on demand empty.
“Err4” requested set parameter is not available.
“Err5” command argument is wrong.
“Err6” requested range not available.
“Err7” meter in log mode.
“Err8” is sent if instrument is not in measurement mode.
“NAK” (21) character is sent when the instrument receives
an unknown or a corrupted command.
Note: P ‑ request for pH range.
M ‑ request for mV range.
E ‑ request for EC range.
T ‑ request for TDS range.
N ‑ request for NaCl range.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI2550 and is the answer not in the manual?