OPERATIONAL GUIDE
Each meter is supplied complete with a 9V battery. Remove the
battery compartment cover on the back of the meter (see page 16).
Install the battery while observing its polarity.
Connect the probe to the meter securely
by aligning the pins with the socket and
pushing the plug in.
Make sure that the meter has been cali
brated before taking any measurements
(see “Calibration” section).
Immerse the conductivity probe into
the sample, with the holes on the shaft
completely submerged.
If possible, use plastic beakers or con
tainers to minimize any EMC interference.
Tap the probe lightly on the bottom of the beaker to remove any
air bubbles which may be trapped inside the PVC sleeve.
Turn the instrument on by pressing the On/Off key.
For HI8633 and HI8734:
Take the temperature of the solution
with a ChecktempC or another accurate
thermometer following LCD indication.
Press and hold down the Temp key to display the temperature and
adjust the temperature knob to that of the solution e.g. 20 °C.
For HI8733:
Adjust the TEMPERATURE COEFFICIENT
knob to 2% to compensate for the
temperature effect of average solutions
(to determine exact value for a particular
solution, see page 15).
Select the appropriate measurement range.