EasyManuals Logo

Hanna Instruments HI8733 User Manual

Hanna Instruments HI8733
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
9
Note: If the display shows only a “1” on
the far left hand side, the meter is out of
range. Select the next (higher) range.
Wait for a couple of minutes for the temperature sensor to reach
thermal equilibrium with the sample before taking measurements.
After the measurement has been completed, the instrument should
be switched off and the probe should be cleaned and dried (see
“Probe Maintenance” on page 17).
CALIBRATION
Accessories needed:
Use any calibration solution within the meter’s range. The
solution should ideally be close to the samples being measured.
Use for example HI7030 or HI8030, 12880 µS/cm (=12.88
mS/cm) conductivity solution, for HI8633 and HI8733, and
HI7032, 1382 mg/L (=2764 µS/cm) TDS solution, for
HI8734.
ChecktempC or another accurate thermometer with 0.1 °C
resolution (not necessary for HI8733).
PROCEDURE FOR HI8633 AND HI8734
Pour sufficient quantity of a conductivity
(HI8633) or TDS (HI8734) calibration
solution (e.g. HI7030 or HI7032) into a
beaker to cover the holes on the probe. If
possible, use plastic beakers to minimize
any EMC interference.
Immerse the conductivity probe, making
sure that holes are completely submerged,
and the ChecktempC in the solution.
Wait for a couple of minutes for thermal
equilibrium to be reached.
Tap the probe on the bottom, then shake it while rotating to make
sure no air bubbles remain trapped in the sleeve.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI8733 and is the answer not in the manual?

Hanna Instruments HI8733 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI8733
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals