No Light:The instrument cannot adjust the light level. Please check
that the sample does not contain any debris.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Zero: A zero reading was not taken. Follow the instructions of the
measurement procedure for zeroing the meter.
b) on sample reading
Under range: A blinking “0” indicates that the sample absorbs
less light than the zero reference. Check the procedure and make
sure you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration
indicates an over range condition. The concentration of the sample
is beyond the programmed range: dilute the sample and re‑run
the test.
ERRORS AND WARNINGS
Standard High: The standard reading is higher than expected.
Standard Low: The standard reading is less than expected.
c) during calibration procedure