GENERAL TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure best accuracy.
Color or suspended matter in large amounts may cause interference, therefore these should be
removed by treatment with active carbon and by prior filtration.
For a correct filling of the cuvette: the liquid in the cuvette forms a concavity on the top; the
bottom of this concavity must be at the same level of the 10 mL mark.
Proper use of the powder reagent packet:
(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.
It is important that the sample does not contain any debris. This would corrupt the reading.
Each time the cuvette is used, the cap must be tightened to the same degree.
Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely
free of fingerprints, oil or dirt. Wipe it thoroughly with HI731318 or a lint‑free cloth prior to
insertion.