20
CALIBRATION PROCEDURE
Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking
“ZERO” asks for instrument zeroing.
Press and hold CAL CHECK for three seconds to enter calibration mode. The display will show
“CAL during calibration procedure. The blinking “ZERO” asks for instrument zeroing.
Place the CAL Check
TM
Standard HI96751‑11 Cuvette A into the cuvette holder and ensure
that the notch on the cap is positioned securely into the groove.
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or
ON/OFF keys.
Warning: Do not calibrate the instrument with standard solutions other than the Hanna Instruments
CAL Check™ Standards, otherwise erroneous results will be obtained.
3''