METHOD PROCEDURE
• Remove the cuvette.
To an empty cuvette add 3 drops of
HI93701A-F Free Chlorine Reagent A and
3 drops of HI93701B-F Free Chlorine
Reagent B.
• Swirl gently to mix.
Add 10 mL of unreacted sample (up to the
mark). Replace the plastic stopper and the
cap. Shake gently to mix.
Insert the cuvette into the holder and ensure
that the notch on the cap is positioned securely
in the groove.
• Press Read. The display will show a 1 minute countdown prior to the measurement.
To skip the timer, press Read twice. When the timer ends, the meter will perform the
reading. The instrument displays the results in mg/L of chlorine (Cl
2
).
×3 ×3
10 mL