26
METHOD PROCEDURE
8.3. TOTAL CHLORINE UHR
REQUIRED REAGENTS
Code Description Quantity
HI95771A-0 Total Chlorine UHR Reagent A 1 packet
HI95771B-0 Total Chlorine UHR Reagent B 1 packet
REAGENT SETS
HI95771-01 Total Chlorine UHR Reagent - 100 tests
HI95771-03 Total Chlorine UHR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the Total Chlorine UHR method using the procedure described in the METHOD
SELECTION section.
Note: If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press Measure and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL