pH CALIBRATION
Press CAL. The instrument will display the measured pH, the LCD first expected buffer and the
temperature reading.
If necessary, press the / keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and within range of the selected buffer, CFM functional key is
displayed.
FIVE POINT CALIBRATION
Immerse the pH electrode approximately 4 cm (1½) into a buffer solution of your choice
(pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 or a custom buffer) and stir gently.
Press CFM to confirm first point.
The calibrated value and the second expected buffer value is then displayed on the LCD.