pH CALIBRATION
If necessary, press the / keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and within range of the selected buffer, the CFM functional key is
displayed.
Press CFM to confirm calibration.
The calibrated value and the fourth expected value will be displayed.
After the third calibration point is confirmed, immerse the pH electrode approximately
4 cm (1½) into the fourth buffer solution and stir gently.
If necessary, press the / keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and within range of the selected buffer, the CFM functional key is
displayed.