36
mV AND TEMPERATURE CALIBRATION (for technical personnel only)
Press CFM to confirm.
The second expected calibrated point is displayed.
Immerse the pH probe including temperature sensor into the second vessel as close as possible
to the reference thermometer. Allow a few seconds for the probe to stabilize.
Use the / keys to set the calibration point value to that of the hot water.
When the reading is stable and within range of the selected calibration point, CFM functional key
is displayed.
Press CFM to confirm. The instrument returns to measurement mode.
Note: Use / keys to change calibration point if necessary (±10.0 °C) around the
point. If the reading is not within range of the selected calibration point, “Wrong”
message will blink. Change pH probe including temperature sensor and restart calibration.