62
Note: If no turbidity calibration has been performed or if calibration was cleared using the
“Restore factory calib.” option the offset and slope values are set to default, and the message
”Factory calibration” is displayed. Press <ESC> to return to the previous screen.
Temperature
• From the “GLP” menu select the “Temperature” option.
• Data regarding the last temperature calibration will be displayed: calibrated point, time and date.
• Use the / keys to scroll through the stored data for the last 5 calibrations.
Notes: If no user temperature calibration has been performed or if calibration was cleared using
the “Restore factory calib.” option the offset value is set to default, and the message ”Factory
calibration” is displayed. Press <ESC> to return to the previous screen.
Atmospheric Pressure
• From the “GLP” menu select “Atm. pressure”.
Data regarding the last atmospheric pressure calibration will be displayed: custom calibration point,
time and date.
• Use the / keys to scroll through the stored data for the last 5 calibrations.
Note: If no atmospheric pressure calibration has been performed or if calibration was
cleared using the “Restore factory calib.” option the offset value is set to default, and the
message ”Factory calibration” is displayed. Press <ESC> to return to the previous screen.
GLP DATA