Chapter 13 - TROUBLESHOOTING/ERROR MESSAGES
HI9829 displays error messages to aid in troubleshooting. Warnings are displayed for most issues,
while Errors are displayed for critical issues.
See the calibration chapter for messages that can occur during calibration. Other messages are listed
below.
“Log space full” appears when the meter memory is full
and additional data cannot be logged or downloaded from
a logging probe. Delete one or more lots from the meter
(Log / Meter Log), or download and delete one or more
logs from the probe.
“Low probe external power supply level! Probe parameters
disabled”: the battery voltage supplied from the meter
to the probe is too low and the measurements could be
adversely affected. All parameters set on probe are disabled.
Press left soft key, check the connection between meter and
probe. If the problem persists, contact your local Hanna
Instruments Office.
“Power fault. Check the probe cable”: this message may
appear when powering up the meter with a probe con
nected. If the meter detects a high load on the probe
connection this message is triggered. Check the probe
cable. If the problem persists, contact your local Hanna
Instruments Office.
“Language data not available”: this message appears when
powering up the meter if the language file is not seen by
the meter. Restart the meter to verify this is a true meter
error. If the problem persists, contact your local Hanna
Instruments Office.
TROUBLESHOOTING/ERROR MESSAGES